Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an uphold¿ lite with capio slim was used a during a procedure.According to the complainant, the dart detached from the suture upon inserting into the capio carrier ("ancillary").Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
 
Manufacturer Narrative
A visual examination of the returned uphold lite revealed that the suture on the blue dilator is broken.The remainder of the suture with dart was not returned.Analysis reveals no damage to the capio slim suture capturing device.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that most probable root cause classification is manufacturing process design because the design or validation of the manufacturing process was not sufficient to ensure the finished device met the intent of the design.The issue is under investigation and a correction has not yet been implemented.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu)/product label.The device was returned with its original packaging, providing the lot number.Blocks now provide the device lot number, expiry and manufacture dates, and blocks have updated manufacturer/manufacturer site information since the device lot is now known.
 
Event Description
It was reported to boston scientific corporation that an uphold lite with capio slim was used a during a procedure.According to the complainant, the dart detached from the suture upon inserting into the capio carrier ("ancillary").Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
MDR Report Key7296224
MDR Text Key100924454
Report Number3005099803-2018-00546
Device Sequence Number1
Product Code OTP
Combination Product (y/n)N
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/19/2020
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0021157076
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Initial Date Manufacturer Received 01/29/2018
Initial Date FDA Received02/26/2018
Supplement Dates Manufacturer Received03/09/2018
Supplement Dates FDA Received03/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-