The system was examined and the reported event was confirmed.The vitrectomy valve cabling was replaced to resolve the issue.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on may 31, 2013.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to a nonconforming vitrectomy valve cabling.The manufacturer internal reference number is: (b)(4).
|