Catalog Number 9734298 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Device manufacturing date is dependent on lot number/serial number, therefore, unavailable no parts have been received by the manufacturer for evaluation.No parts returned for analysis.
|
|
Event Description
|
A medtronic representative reported that while outside of a procedure the 4.5 cannulated tap was found in sterile processing with a bone jammed into the tap that could not be removed.There was no patient present when this issue was observed.
|
|
Manufacturer Narrative
|
Additional information: lot number and manufacture date provided.Correction: product, unique device identification (udi) and associated fields updated to proper value.The tap was returned to the manufacturer for evaluation.It was reported that the cannula of the tap was blocked.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|