The investigation determined that a non-reproducible, higher than expected, vitros tsh result was obtained from a single patient sample when tested on a vitros 5600 integrated system.A definitive assignable cause could not be determined.Based on historical quality control results, a vitros tsh lot 5505 performance issue is not a likely contributor to the event.Additionally, there was no evidence to suggest a vitros 5600 integrated system malfunction; however, it cannot be entirely ruled out as a contributing factor as precision testing to assess instrument performance was not performed by the customer.In addition, pre-analytical sample processing could not be ruled out as a contributing factor, as it was established the customer was not following the sample collection device manufacture¿s recommendation for sample centrifugation.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.
|
A customer obtained a non-reproducible higher than expected vitros tsh result when processing a single patient sample on a vitros 5600 integrated system.Patient sample 1 result of >100 miu/l vs.The expected result of 0.021 miu/l.A biased result of the direction and magnitude observed may lead to inappropriate medical action if not detected.The non-reproducible, higher than expected, vitros tsh result was not reported from the laboratory and ortho was not made aware of any allegation of patient harm as a result of this event.(b)(4).
|