Report event: sparking was reported at the outlet.Device inspection: per gsp case #: 164319 mps dylan smith reported sparking at outlet.- power cord was replaced and problem was resolved.- system successfully passed all validation testing.- verified system is operating within mako tolerances and specifications.- system is ready for clinical use.Device history review: a review of the dhr associated with rob118 found qips passed with no notes or comments.Complaint history: a review of complaints in catsweb and trackwise related to the power cord, tgs system v2, 20' lg, catalog #: 207010, robot #: rob118 shows no additional complaints related to the failure in this investigation.Conclusion: per fse: the power cord was replaced.The system successfully passed all validation testing and is operating within mako tolerances and specifications.The system is ready for clinical use.Further action: none at this time.
|