A radiographic evaluation confirmed the alleged event.The root cause has not been determined as product is pending to be returned.Labeling review: ".Preoperative warnings only patients that meet the criteria described in the indications should be selected.Patient condition and/or predispositions such as those addressed in the aforementioned contraindications should be avoided." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".All set screws should be final-tightened with the counter- torque and torque t-handle.Do not final-tighten through compression instruments in the set, as the rod may not be able to normalize to the tulip." ".All lock screws should be final-tightened with the counter-torque and torque t-handle.Do not final-tighten through compression instruments (e.G.C/d rack and figure 8 compressor) in the set, as the rod may not be able to normalize to the tulip.Be cautious not to over compress or distract as you can loosen the screws in the spine and potentially pull out the screw.The bulleted portion of the nose of the rod and the faceted portion of the rod (where the inserter locks down on the rod) must extend fully outside of the most inferior or most superior tulip on the construct.The set screw cannot be locked down on this unusable portion of the rod, as this may compromise the stability of the construct." ".Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." ".Care should be taken to insure that all components are ideally fixated prior to closure.All implants should be used only with the appropriately designated instrument (reference surgical technique)." ".Potential risks identified with the use of this system, which may require additional surgery, include: · bending, fracture or loosening of implant component(s) · loss of fixation".
|
On (b)(6) 2017 a patient underwent an excision of lumbar vertebrae and posterior fusion.Approximately two months post-operative two lock screws backed out, separated from pedicle screws and protruded through the skin.On (b)(6) 2018 a revision procedure was performed replacing the pedicle screws, lock screws and rods without no reported issues.
|