Model Number N/A
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.Device has not yet been returned for evaluation.
 
Event Description
Per tm the sherlock (b)(4) is showing the picc coming in from the opposite side they are working from in addition to dropping off the screen.No other information was provided.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of inaccurate magnet tracking was unconfirmed.The magnet tracking and ecg readings were correct as received.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to refurbished inventory.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per tm the sherlock (b)(4) is showing the picc coming in from the opposite side they are working from in addition to dropping off the screen.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURB, SHERLOCK 3CG TCS
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7308230
MDR Text Key101478210
Report Number3006260740-2018-00321
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741124341
UDI-Public(01)00801741124341
Combination Product (y/n)N
PMA/PMN Number
K113808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number9770131
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2018
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/01/2018
Supplement Dates Manufacturer Received08/28/2019
Supplement Dates FDA Received08/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-