BARD ACCESS SYSTEMS REFURB, SHERLOCK 3CG TCS; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.Device has not yet been returned for evaluation.
|
|
Event Description
|
Per tm the sherlock (b)(4) is showing the picc coming in from the opposite side they are working from in addition to dropping off the screen.No other information was provided.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of inaccurate magnet tracking was unconfirmed.The magnet tracking and ecg readings were correct as received.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to refurbished inventory.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per tm the sherlock (b)(4) is showing the picc coming in from the opposite side they are working from in addition to dropping off the screen.No other information was provided.
|
|