The lithocrush device was reported discarded by the user facility and will not return to olympus for evaluation.The cause of the customer complaint cannot be confirmed.A potential cause for the dislodged tip is excess applied force.The instruction manual contains warnings and cautions to prevent mechanical damage: ¿when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire¿be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire¿this may damage the distal tip¿.¿never use excessive force to operate the instrument and bml handle.¿ and ¿insert the instrument slowly.¿ the instruction manual also has directions for pre-procedure visual inspection and operational exercise of the device.As a preventive measure in the event of device malfunction, the instruction manual also advises to always have a spare device available during the procedure.
|