Model Number 91390 |
Device Problems
Loss of Power (1475); Failure to Power Up (1476)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Spacelabs has launched an investigation and will file a supplemental report once the investigation is complete.
|
|
Event Description
|
Spacelabs received a report that on (b)(6) 2018, while on a patient, the qube monitor shut off and would not reboot.No injury was reported with this event.
|
|
Manufacturer Narrative
|
The qube compact monitor was received at spacelabs¿ equipment service center for repair where the reported problem was verified.The printed circuit board assembly with power supply was replaced.The repaired unit passed all functional tests and was returned to the customer.This report is complete and this particular issue is considered closed.
|
|
Search Alerts/Recalls
|