Product event summary: the data files and balloon catheter, 2af284 with lot number 56485 were returned and analyzed.The data files showed at least thirty-six injections were performed with two catheters for the date of the event.System notice 50005 ¿the safety system has detected fluid in the catheter and stopped the injection¿ was seen in the failure files.Visual inspection of the catheter showed blood residue inside the balloon.The catheter failed the performance test due to the 50005-system notice.Dissection/ pressure testing showed a guide wire lumen was kinked/ twisted at 1.61 inches proximal from the tip and revealed a guide wire lumen breach at 1.61 inches from the tip.In conclusion, the reported issue was confirmed through testing and data analysis.The balloon catheter failed the inspection due to a guide wire lumen kink and breach.If information is provided in the future, a supplemental report will be issued.
|