• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION VALVE, O.P.S (STERILE) FOR TC; OVERPRESSURE SAFETY VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION VALVE, O.P.S (STERILE) FOR TC; OVERPRESSURE SAFETY VALVE Back to Search Results
Model Number LH130J
Device Problems Defective Device (2588); Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 01/16/2018
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during non - clinical activity, the duckbill valve was defective.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 5, 2018.(b)(4).Visual inspection was performed on the returned sample, during which no anomalies were noted anywhere on the device.Ops valves are subject to a 100% leak test, which includes five different tests the units are run through to ensure there are no leaks and both the umbrellas and duckbills are functioning properly.Upon evaluation of the returned sample, it was found to function as intended, and met all of the product specifications.Based on the complaint description, the suction or flow through the ops valve was not able to meet the specification during an inspection at distributor facility.This is believed to be an issue with the forward flow through the duckbill valve.The reported issue was not able to be replicated; therefore, the complaint was not confirmed.All ops valves undergo 100% leak test and visual inspection.There is a known issue currently ongoing with the duckbill valve, within the ops valve, that prevents the duckbill from opening within the product specification range.It is possible that the duckbill valve did not open during the event, as reported; however, if a high enough pressure had been reached after that initial attempt, the seal in the slit of the duckbill may have been broken, allowing the valve to pass all leak tests during the evaluation of the returned sample.This was not able to be confirmed.The cause of the issue with the duckbill is due to a process change at the supplier, specifically during the washing/cleaning step.The new validated process has already been implemented at the supplier and terumo has initiated an action for removal of all affected units from the field.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE, O.P.S (STERILE) FOR TC
Type of Device
OVERPRESSURE SAFETY VALVE
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key7313728
MDR Text Key101519342
Report Number1124841-2018-00034
Device Sequence Number1
Product Code DTL
Combination Product (y/n)N
PMA/PMN Number
K820297
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2020
Device Model NumberLH130J
Device Catalogue NumberN/A
Device Lot NumberVF01
Other Device ID Number(01)00699753450233
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2018
Initial Date Manufacturer Received 02/13/2018
Initial Date FDA Received03/05/2018
Supplement Dates Manufacturer Received04/04/2018
Supplement Dates FDA Received04/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number1124841-06/25/2017-001-R
Patient Sequence Number1
-
-