Manufacturing evaluation: the device was not available for return.Failure analysis without product: according to the instructions for use (ifu), the device is packaged "nonsterile- reusable".The ifu specifically states under how supplied: aesculap prestige endoscopic graspers are provided non-sterile, packaged one per box.Each instrument must be sterilized before use according to the procedures described in the sterilization section of this ifu document.Section 8, sterilization, page 10 of the ifu, gives instructions on how to sterilize the device including minimum exposure time and temperature.Neither micro nor aesculap made any claims that the device would be sterile upon receipt by the end user.If it was after sterilization that bioburden was found then the end customer did not follow proper instructions for sterilization.
|