ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 36CM; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HAR36
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/16/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p92n4e.Investigation summary the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.In addition no damaged at the tissue pad was observed at any time during testing was reported by the user.This blade tip portion may have broken off of the device during transport to our analysis site.The reported complaint was that the tissue pad was loosed.During functional testing on gen11 an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during an unknown procedure, the tissue pad was loosed of the material.It was necessary to use another material.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC LAP 5MM SHEAR 36CM
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7314589
MDR Text Key101712658
Report Number3005075853-2018-08359
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036014444
UDI-Public20705036014444
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K120729
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 02/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue NumberHAR36
Device Lot NumberP92N4E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2018
Initial Date Manufacturer Received 02/15/2018
Initial Date FDA Received03/05/2018
Date Device Manufactured07/25/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-