• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL, INC AVEA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC AVEA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number AVEA
Device Problems Occlusion Within Device (1423); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 02/14/2018
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).Any additional information received from the customer will be included in a follow-up report.At this time, the vyaire failure analysis laboratory has received the suspect gas delivery engine (gde) but it is still under investigation.The completed investigation will be included in a follow-up report.
 
Event Description
The customer reported an unresolved extended high peak and circuit occlusion alarm, on this ventilator device.The customer reported that there was no patient involvement.
 
Manufacturer Narrative
The vyaire failure analysis laboratory received the suspect transducer communication alarm (tca) assembly for investigation.The physical inspection found no anomalies with the tca.The investigation did not duplicate the customer event during the performance and manufacture testing.The alarm conditions and device behavior were not present during the component testing of the transducer communication alarm (tca) assembly.A component cause of this failure was not determined during the investigation therfore, no component root cause could be determined.The tca assembly will be held for trending and internally investigated within vyaire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEA VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7314769
MDR Text Key101566589
Report Number2021710-2018-07479
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103211
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAVEA
Device Catalogue Number17310-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2018
Initial Date FDA Received03/05/2018
Supplement Dates Manufacturer Received08/06/2018
Supplement Dates FDA Received08/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2008
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-