• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) DYNAMIC XT¿ UNIDIRECTIONAL STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) DYNAMIC XT¿ UNIDIRECTIONAL STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number M0042011010
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiac Arrest (1762); Cardiogenic Shock (2262)
Event Date 02/13/2018
Event Type  Injury  
Manufacturer Narrative
Device evaluated by mfr: it is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.The investigation conclusion is device not returned as the complaint did not include returned device review and lacked the objective evidence or descriptive conditions of the event required to determine what could have contributed to the event.(b)(4).
 
Event Description
Same case as mdr id# 2134265-2018-01862, 2134265-2018-01863, and 2134265-2018-01976.It was reported cardiac arrest and cardiogenic shock occurred.Two intellamap orion mapping catheters, a dynamic xt steerable diagnostic catheter, and a woven diagnostic catheter were selected for use during a ventricular tachycardia ablation procedure.During ablation with another manufacturer¿s ablation catheter, the patient experienced cardiac arrest and cardiogenic shock.Multiple external defibrillations/shocks were administered, as well as advanced life support including cpr and medications.The patient ultimately received circulatory assist devices (a heart pump and extracorporeal membrane oxygenation (ecmo)) before being transferred from the ep lab.The ep staff later reported that the patient required vascular surgery the next day as a result of the ecmo device.The patient appeared to be stabilized but in critical condition.Noise with the orion catheters had been noted; however, there was no information alleging the noise caused or contributed to the complications.The orion catheters had been removed from the patient¿s body for two and a half hours prior to onset of the patient¿s complications.The dynamic xt was in the patient when the complications occurred, and the woven catheter was positioned in the right ventricle thorough out the procedure.The physician was uncertain as to the cause of the cardiac arrest and cariogenic shock, but commented that the patient was likely sicker than previously considered prior to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAMIC XT¿ UNIDIRECTIONAL STEERABLE DIAGNOSTIC CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7316051
MDR Text Key101582294
Report Number2134265-2018-01864
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K921872A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/04/2021
Device Model NumberM0042011010
Device Catalogue Number201101
Device Lot Number21576010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/13/2018
Initial Date FDA Received03/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ABLATION CATHETER: BIOSENSE WEBSTER SF
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age66 YR
-
-