(b)(6).After spontaneous movement a crackling sound in the right side of hip was noticed by the patient.Fracture of ceramic head could be detected by x-ray.Revision surgery with removing the broken head and changing the inlay was on (b)(6) 2018.Primary surgery was on (b)(6) 2001 at (b)(6).Components in use listed as concomitant devices are: nk460 / biolox prosthesis head 12/14 28mm s, nh054t / plasmacup sc size 54mm, nh093 / sc/msc ceramics insert 28mm 52/54 sym., nk662t / antega plasmapore 12/14 size 12mm right.
|
Investigation: the ceramic ball head cannot be completely reconstructed from the delivered fragments.The density of the fragments was tested and found to be within specification.There are metal transfer patters on the surface of the ball head.Batch history review: the manufacturing history records have been checked and found to be according to specification, valid at the time of production.Conclusion and root cause: on the basis of the investigation, a product related failure can be excluded.There are no indications of a material defect or manufacturing error.The failure is most probably patient (overload) or user (implantation situation) related.No capa is necessary.
|