During preparation for a medical procedure, the hospital staff accidentally dropped and contaminated the indigo system cat3 aspiration catheter (cat3) after removal from the packaging.Therefore, the cat3 was not used in the procedure.The procedure was completed using another catheter.
|
Please note the device is no longer available for return as mentioned in the initial mdr; therefore, the mdr was updated accordingly.The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital discarded the device.
|