MEDOS INTERNATIONAL SARL EXPRESSEW III AUTOCAPTURE+ SUTURE PASSER; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 288233
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/16/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
It was reported by the sales rep that after a rotator cuff repair, the top jaw on the customer's expressew auto capture gun would not close all the way and seemed to be bent.The surgeon completed the surgery with the device with no patient consequences or delays.The device is being returned.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device is not being returned, multiples attempts for product return were made with no response, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A device history record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III AUTOCAPTURE+ SUTURE PASSER
Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key7320276
MDR Text Key101968182
Report Number1221934-2018-50278
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705027101
UDI-Public(01)10886705027101(10)41488-170210
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number288233
Device Catalogue Number288233
Device Lot Number41488-170210
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/16/2018
Initial Date FDA Received03/07/2018
Supplement Dates Manufacturer Received11/15/2018
Supplement Dates FDA Received11/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-