Model Number 560BC |
Device Problems
Device Operates Differently Than Expected (2913); Pumping Problem (3016)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/22/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Product has not been returned for analysis.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of the bio-console 560 there was a burning smell from the console.The instrument was replaced with a backup.Impact to patient was asked but unknown.Additional information was received that the cardiopulmonary bypass was temporarily performed by manual means using the hand crank while the bio-console was replaced.
|
|
Manufacturer Narrative
|
The bio-console instrument was returned and the reported burning smell was verified during service.Medtronic service noted that the diode on the motor-control board was burned and damaged.The issue was resolved by replacing the motor-control board.During service the sc board was also replaced due to an error with resolve.The impedance of the rpm pot was stabilised.A nut for the fan was detached, the issue was resolved by repairing the nut.The internal battery and fan filer were replaced as part of preventive maintenance.Functional testing was performed per specifications.Trends for issues with this product are monitored.If information is provided in the future, a supplemental report will be issued.
|
|