Model Number 8070
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Conclusion: evaluation of the returned product found one element missing on the molding zipper (a patient access area) which could allow the unintentional release of a patient.Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper.In addition, there was a one inch tear on the head side leg and multiple areas of unraveling on the head and foot side legs.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing.Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.(b)(4).
 
Event Description
Customer reported the zipper on the left side panel is damaged.The date the issue was discovered is unknown.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8070
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
POSEY PRODUCTS LLC
5635 peck road
arcadia CA 91006
Manufacturer (Section G)
POSEY S DE RL DE CV
cam. antigua a tecate #16760 i
colonia niños heroes este
tijuana, baja california 22120
MX   22120
Manufacturer Contact
william hincy
posey products llc
5635 peck road
arcadia, CA 91006
6264433143
MDR Report Key7327787
MDR Text Key102201072
Report Number2020362-2018-00019
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113357
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Model Number8070
Device Catalogue Number8070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/07/2018
Initial Date FDA Received03/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-