Model Number 8070
Device Problem Component Missing (2306)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the returned canopy found one element missing on the right window drainage port and one element missing on the molding zipper a patient access area.If one zipper element is missing or pulled apart on a patient accessible zipper it would allow the zipper to be initially closed.A patient can later open the zipper by separating the zipper at the location of the missing element.Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing.Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.(b)(4).
 
Event Description
Customer reported the zipper teeth are missing from the right side panel of the canopy.The date the issue was discovered is unknown.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8070
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
POSEY PRODUCTS LLC
5635 peck road
arcadia CA 91006
Manufacturer (Section G)
POSEY S DE RL DE CV
cam. antigua a tecate #16760 i
colonia niños heroes este
tijuana, baja california 22120
MX   22120
Manufacturer Contact
william hincy
posey products llc
5635 peck road
arcadia, CA 91006
6264433143
MDR Report Key7327795
MDR Text Key102215117
Report Number2020362-2018-00020
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103817
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Model Number8070
Device Catalogue Number8070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/19/2018
Initial Date FDA Received03/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-