MEDTRONIC NAVIGATION, INC. (LOUISVILLE) INTERBODY INSERTER CAPSTONE & CLYDESDALE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734456
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/15/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.No parts have been received by the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that during spinal fusion procedure, the tip of the inserter bent upon insertion.The procedure was completed with the use of navigation.There was no delay to procedure.No impact on patient outcome.
 
Manufacturer Narrative
Correction: instrument unique device identification (udi) updated to proper value.
 
Manufacturer Narrative
Correction: initial reporter: (b)(6).
 
Manufacturer Narrative
The capstone and clydesdale inserter was returned to the manufacturer for analysis.Analysis found that the tip of the inserter was broken off on one side.The cause of the breakage could not be determined.Analysis found that the reported event was related to a physical damage issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERBODY INSERTER CAPSTONE & CLYDESDALE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7331072
MDR Text Key102139056
Report Number1723170-2018-01104
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup,Followup
Report Date 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734456
Device Lot Number120613
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/15/2018
Initial Date FDA Received03/12/2018
Supplement Dates Manufacturer Received03/21/2018
09/11/2018
09/14/2018
Supplement Dates FDA Received04/11/2018
09/11/2018
09/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/13/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-