• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC 1818910 SELF CENT HIP 40X28 RUST; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC 1818910 SELF CENT HIP 40X28 RUST; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Catalog Number 103540000
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/16/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Der states that the bipolar liner would not lock into bipolar shell.Surgical delay- 5 minutes; doe: (b)(6) 2018.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Der states that the bipolar liner would not lock into bipolar shell.Surgical delay- 5 minutes; doe: (b)(4) 2018.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # pc-(b)(4).Investigation summary examination of the returned device is not able to confirm the reported problem.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 40X28 RUST
Type of Device
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS INC 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7331274
MDR Text Key102119084
Report Number1818910-2018-54882
Device Sequence Number1
Product Code KWY
UDI-Device Identifier10603295003304
UDI-Public10603295003304
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K812672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number103540000
Device Lot NumberHG8431
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2018
Initial Date FDA Received03/12/2018
Supplement Dates Manufacturer Received04/10/2018
05/29/2018
11/08/2018
Supplement Dates FDA Received05/02/2018
05/30/2018
11/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/25/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient Weight72
-
-