Model Number CS08657-01 |
Device Problem
Smoking (1585)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product is not sold in us and is 510(k) exempt.This report is associated to a similar product sold in the us - 510(k) number k892432.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, there was a black smoke appears in the unit when it was plugged.The customer indicated that there was no patient involvement.
|
|
Manufacturer Narrative
|
Correction: (udi).Evaluation summary one sample was received for evaluation, a visual inspection was conducted and it was observed that the ferrite core pieces were found from damaged component l1 on the ac/dc pcba.Heat sink screw for u7 had small black marks and damage consistent with arcing.Ferrite core pieces were inspected and discoloration consistent with heat was found.Dc/dc pcba was removed.Potential root cause determined to be broken ferrite pieces causing arcing between the dc/dc pcba u7 heat sink screw and the chassis.If information is provided in the future, a supplemental report will be issued.
|
|