• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® SUCTIONAID® TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® SUCTIONAID® TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/515/075
Device Problem Inflation Problem (1310)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/05/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).This device is same and/or similar to a device approved for distribution in the us.
 
Event Description
It was reported that the cuff of a portex® suctionaid® tracheostomy tube inflated normally during pre-use check, but not after it was intubated in the patient.No injury was reported.
 
Manufacturer Narrative
Two used samples were received for investigation.A visual inspection could not find any holes.Use testing was able to confirm to the reported problem as leakage was detected in the cuffs of both samples.A manufacturing review of a similar device was conducted and no discrepancies were found.A root cause for the defect could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SUCTIONAID® TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
7633833310
MDR Report Key7332051
MDR Text Key102205688
Report Number3012307300-2018-00397
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date01/06/2020
Device Catalogue Number100/515/075
Device Lot Number3521444
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2018
Initial Date FDA Received03/12/2018
Supplement Dates Manufacturer Received05/04/2018
Supplement Dates FDA Received06/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-