According to the complaint, a customer transferred blood from a syringe to two separate capillary tubes and ran the blood samples on an abl835 analyzer and on an abl90 flex plus analyzer.The customer reported that there were a significant difference between results from the two analyzers on the pco2, thb, k and glucose results.The customer obtained the following results: (b)(6).The customer stated that the results from the abl835 analyzer more accurately reflected the patient's condition.Later on the same day, a field service engineer visited the site and ran two blood samples on the same analyzers although the results were not as discrepant as the results obtained by the customer.
|