STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2
Device Problems Improper or Incorrect Procedure or Method (2017); Device Operates Differently Than Expected (2913)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.Work order search: no similar complaint was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated the surgeon was unable to inject a 13.2mm vticm5_13.2 implantable collamer lens, - 10.0/+2.0/091 (sphere/cylinder/axis), due to wrong viscoelastic was used.The lens was damaged during the loading process.The lens was not implanted and there was no report of patient contact.The backup lens was implanted.
 
Manufacturer Narrative
Device evaluation: the lens was returned dry on a micro centrifuge vial with presence of clear residue on the product.Visual inspection found that the lens was missing a part and the presence of residue on lens surface.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key7339933
MDR Text Key102468419
Report Number2023826-2018-00415
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date05/31/2020
Device Model NumberVTICM5_13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/12/2018
Initial Date FDA Received03/14/2018
Supplement Dates Manufacturer Received04/05/2018
Supplement Dates FDA Received04/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-