• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-60
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/14/2018
Event Type  malfunction  
Manufacturer Narrative
Serial number is unknown.This information will be provided in a supplemental report if made available.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova (b)(4)manufactures the centrifugal pump 5(cp5).The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a incorrect calibration of the device.The technician performed a zero calibration to resolve the reported malfunction.Subsequent functional verification testing was completed without further issues and the unit was returned to service.It was not possible to determine why the values were out of specification.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device evaulated by livanova rep.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump 5 (cp5) could not be used properly during procedure.The customer reported that if the knob of the drive unit was used to adjust the speed the unit worked properly but if the knob was released the drive unit stopped.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7341096
MDR Text Key102729098
Report Number9611109-2018-00227
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2018
Initial Date FDA Received03/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-