• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC STEM TRIAL EXTRACT; HIP INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC STEM TRIAL EXTRACT; HIP INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Catalog Number 865226
Device Problem Break (1069)
Patient Problem Not Applicable (3189)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the stem broken and stuck in tibia 14x115, removal tool broke off when trying to retrieve.No pieces of broken instrumentation left in patient.Surgical delay approximately 30 minutes.
 
Manufacturer Narrative
Product complaint: pc-(b)(4).Investigation summary:examination of the returned device confirmed the reported event.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC STEM TRIAL EXTRACT
Type of Device
HIP INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7341107
MDR Text Key102465192
Report Number1818910-2018-55156
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295216766
UDI-Public10603295216766
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number865226
Device Lot NumberG0409
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2018
Initial Date FDA Received03/15/2018
Supplement Dates Manufacturer Received02/28/2018
04/04/2018
Supplement Dates FDA Received03/19/2018
04/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-