Model Number 16402
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Date 02/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The fsr removed the broken tongue and installed a new one.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the tongue on the roller pump was broken.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.The product surveillance technician (pst) visually verified that the tongue was broken.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 UNIVERSAL ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE¿SARNS 8000 UNIVERSAL ROLLER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7342041
MDR Text Key102597942
Report Number1828100-2018-00134
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/21/2018
Initial Date FDA Received03/15/2018
Supplement Dates Manufacturer Received03/29/2018
Supplement Dates FDA Received04/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-