• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5_12.6
Device Problem Activation Failure (3270)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2018
Event Type  malfunction  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that after the surgeon implanted a 12.6mm vicm5_12.6 implantable collamer lens, -5.5 diopter into the patient's right eye (od), the lens did not unfold.Reportedly, the folded lens was inserted into the anterior chamber and it did not open but remained folded.This occurred on (b)(6) 2018.Intraoperatively the lens was exchanged with a replacement lens.The problem is resolved.
 
Manufacturer Narrative
Device evaluation: the lens was returned dry, in a micro centrifuge vial.There was residue on product.Visual inspection found the haptic bent and foreign material on lens surface (residue on lens).(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
8002927902
MDR Report Key7356132
MDR Text Key103079467
Report Number2023826-2018-00429
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date08/31/2020
Device Model NumberVICM5_12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/22/2018
Initial Date FDA Received03/20/2018
Supplement Dates Manufacturer Received04/04/2018
Supplement Dates FDA Received04/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
-
-