The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
|
During preparation for a medical procedure, the hospital staff inadvertently kinked the indigo system cat5 aspiration catheter (cat5) on the back table while taking it out of the hoop.The damage to the cat5 occurred prior to use.Therefore, the cat5 was not used in the procedure.The procedure was completed using a new cat5.
|