Model Number QB55551002 |
Device Problems
Component Falling (1105); Material Frayed (1262)
|
Patient Problems
No Information (3190); No Code Available (3191)
|
Event Type
Injury
|
Manufacturer Narrative
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
|
Event Description
|
According to the available information, when using the keith needle on the dacron pull suture - when you deploy it through the glans the dacron suture is frayed and fall to bits.This means you cannot locate the implant in the shaft.There appears to be no clinical consequences.The operation time was excessive, but coloplast has not been alerted to any infection or impaired function.
|
|
Manufacturer Narrative
|
This follow-up mdr is created to document the evaluation of the returned device.A search was completed for other complaints, and there were no other complaints found for this lot number.Visual examination of the returned device revealed clean edges without noticable burs.The device was reviewed following the instructions for use, and the device operated smoothly and did not result in damage to the suture.A lot review was completed, which didn't reveal any recorded anomaly in connection with the described event.Based on the above, this complaint is not confirmed as a product defect.
|
|