• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS CORPORATION - MONTGOMERY ARMBOARD WITH GRAVITY LATCH; ARMBOARD TO SURGICAL TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CORPORATION - MONTGOMERY ARMBOARD WITH GRAVITY LATCH; ARMBOARD TO SURGICAL TABLE Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Injury (2348)
Event Date 02/22/2018
Event Type  malfunction  
Manufacturer Narrative
The armboard was removed from service following the event.Based on the description of the event; it appears the armboard had been damaged prior to use.The armboard instructions for use state: warning: personal injury and/or equipment damage hazard- do not use equipment if worn, damaged or cannot be securely tightened.Cracked or splintered surfaces may cause injury.Inspect armboard prior to use.Damaged armboard must be replaced.Read and understand all instructions prior to use." in addition, every armboard has a warning label which states: warning-avoid patient or user injury.Check accessory for damage and wear prior to use.Do not use if any damage is apparent or device cannot be securely tightened.The steris account manager completed in-service training on 3/21/2018 with user facility personnel on the importance of inspecting the armboard prior to use.
 
Event Description
The user facility reported that an employee was injured on the carbon fiber surface of an armboard while reaching under the armboard to release it from the surgical table.No medical treatment was sought or administered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMBOARD WITH GRAVITY LATCH
Type of Device
ARMBOARD TO SURGICAL TABLE
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key7363500
MDR Text Key103413766
Report Number1043572-2018-00026
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/23/2018
Initial Date FDA Received03/22/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-