Customer reported during rbc procedure the replacement fluid line is taking the fluid in a reverse way (from the air detector chamber to replacement fluid bag).A technician went to the customer and checked the pl1 set.After close and accurate checking it was observed that the white pump line segment was glued in a reverse way so that instead of withdrawing the replacement fluid from replacement fluid bag to the air detector chamber, it is taking the fluid from air detector chamber and inserting it into the replacement fluid bag.The manufacturer received 4 pictures for evaluation.On the pictures received, it is visible that the tubing assembled to the transparent pump have been exchanged.The assignable cause was determined to be an assembly failure during manufacturing.
|