U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR 2 CANNULA; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR 2 CANNULA; CAT Back to Search Results
Model Number OJR416
Device Problem Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The optiflow junior cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint ojr416 junior cannula was received and was visually inspected.Results: visual inspection of the returned cannula showed that the tube had detached from the cannula.Evidence of adhesive was present on the external surface of the detached tube.Conclusion: the observed damage was most likely caused by an excessive pulling force.All optiflow junior cannulae are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, a tube tensile strength test is carried out and if the tube breaks before the load of 10 newtons is reached the entire batch is scrapped.Samples are also taken from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 cannula.They also state the following: - do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).
 
Event Description
A hospital in (b)(6) reported that the tubing on one side of an ojr416 optiflow junior 2 nasal cannula had become separated from the interface while in patient use.It was confirmed that there was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR 2 CANNULA
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
sharon kuruvilla
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key7364602
MDR Text Key103672915
Report Number9611451-2018-00219
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeNZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOJR416
Device Catalogue NumberOJR416
Device Lot Number2100304809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2018
Initial Date FDA Received03/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-