(b)(4).The device was returned to the factory for evaluation.A visual inspection was conducted.Signs of clinical use and evidence of blood were observed.Traces of blood and charred tissues were observed on the jaws.Microscopic inspection showed the heater wire to be flexed away but remained attached at the tip and at the base of the jaw.The silicone insulation appeared intact with no cracks nor peeling.A pre-cautery test was performed per the instruction for use (ifu) with a reference cable, adapter and reference power supply at level 3.0.During the pre-cautery test, it produced steam and audible beeping sound during five (5) 3-second activations.To evaluate the safety shut down system, a poly-fuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after a 10-second cooling period with no incident each time.An activation and transection capability test was performed over five (5) repetitions using max life test method /bacon test.The jaws were cleaned per the ifu and the pre-cautery test repeated.The device activated and transected tissue six (6) times with no cautery failure observed.No smoke and/or steam which could be considered excessive were observed.A temperature and resistance evaluation was conducted to evaluate the device function.The resistance value was measured at 0.665 ohms which is within hemopro 2 final test specification.The device passed the temperature measurement test.The displayed temperature increased and turned ¿green¿ within the 2 second specified timeframe.The displayed temperature decreased once the toggle switch was released.Based on the returned condition of the device and the results of the investigation, the reported failure modes "failure to cut" and "melted" were not confirmed.The reported failure mode "bent wire" was confirmed.Specific actions for the reported failure mode "bent wire" are being maintained and documented under maquet¿s failure investigation report (fir) system.
|