During preventive maintenance, the autopulse platform (sn (b)(4)) displayed an error message system error, out of service, revert to manual cpr".The defective processor board was replaced to remedy the fault.The autopulse platform was manufactured in 2006 and is 12 years old, well beyond the expected service life of five years.The platform failed the initial functional testing due to the error message "system error, out of service, revert to manual cpr" displayed upon powering on the device.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaints reported for autopulse platform sn (b)(4).
|
The autopulse platform (sn (b)(4)) was returned for preventive maintenance and as part of routine service during testing, the platform was examined and displayed a system error, out of service, revert to manual cpr message on the user control panel during initial power up.
|