• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP Back to Search Results
Model Number 801041
Device Problems Leak/Splash (1354); Pumping Stopped (1503)
Patient Problem No Patient Involvement (2645)
Event Date 02/14/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Per the manufacturer's subsidiary, the belt looses friction and just spins on the shaft stopping the pump.This is a demonstration unit only and not used clinically.
 
Event Description
It was reported that during routine testing of the device at the manufacturing engineering center (mec), grease had leaked from the bearing and the pump stopped.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.Multiple diligence attempts by the manufacturer were unsuccessful in obtaining the part for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7369959
MDR Text Key103568277
Report Number1828100-2018-00151
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/01/2018
Initial Date FDA Received03/26/2018
Supplement Dates Manufacturer Received04/27/2018
Supplement Dates FDA Received04/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-