Device evaluated by mfr: returned product consisted of a ffr comet pressure guidewire.The rfid cable was not returned.The tip, device shaft, sensor port and the modulation values were examined for damage or any irregularities.The guidewire shaft showed 2 kinks.The 1st kink was located at 126cm from the tip and the 2nd kink at 139.5cm from the tip.There also was some peeling of the coating at these locations.No other damage was noticed on the shaft.The sensor port was clear.Since the rfid cable was not returned, a test cable was used to check for modulation.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire.The coefficient values could not be confirmed due to the devices rfid cable not returning with the device.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
|