Model Number 7209208 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the black lens/camera attachment piece became un-attached from the rest of the scope when it was being cleaned.It is very apparent when anyone looks at the scope.There was no patient involvement.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Evaluation summary: one device sample was received for evaluation.The returned sample did not meet specification as received.The customer reported that the black lens/camera attachment piece became un-attached from the rest of the scope when it was being cleaned.The reported condition was confirmed.An evaluation was performed and showed the scope to have scratched fibers and the scope needs cosmetic refurbishing.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|