Additional information: the actual device was not available; however, photographs of the sample were provided for evaluation.The visual inspection of the provided photographs revealed that the pbp and effluent lines were incorrectly assembled on the support plate (pre-pump pbp line inverted with post-pump effluent line).The reported condition was verified.The cause of the condition was determined to be an assembly issue during the manufacturing process.To address this issue, the involved operator has been retrained and awareness training was provided to all operators.The reported lot was manufactured prior to this action.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|