• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX HF1400; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX HF1400; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 107142
Device Problem Misassembled (1398)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/26/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the effluent line (yellow) was mismatched with the pre blood line (white) line of a prismaflex hf1400 set.This issue was identified when the baxter crrt machine alarmed almost immediately after the patient was connected.Troubleshooting revealed that the effluent and pbp lines had been misassembled in the hf1400 filter set.Treatment was terminated without returning the extracorporeal blood to the patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: the actual device was not available; however, photographs of the sample were provided for evaluation.The visual inspection of the provided photographs revealed that the pbp and effluent lines were incorrectly assembled on the support plate (pre-pump pbp line inverted with post-pump effluent line).The reported condition was verified.The cause of the condition was determined to be an assembly issue during the manufacturing process.To address this issue, the involved operator has been retrained and awareness training was provided to all operators.The reported lot was manufactured prior to this action.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX HF1400
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu cedex rhone
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
ni
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7372402
MDR Text Key103555852
Report Number8010182-2018-00018
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414069315
UDI-Public(01)07332414069315
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042938
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/26/2018,04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/01/2019
Device Catalogue Number107142
Device Lot Number17I0704
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/26/2018
Distributor Facility Aware Date02/26/2018
Event Location Outpatient Treatment Facility
Date Report to Manufacturer03/26/2018
Initial Date Manufacturer Received 02/26/2018
Initial Date FDA Received03/26/2018
Supplement Dates Manufacturer Received04/12/2018
Supplement Dates FDA Received04/19/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-