A review of the device history records found no manufacturing, processing or design related irregularities.The implants were found to be properly manufactured and released in accordance with the device master record.An evaluation of the returned implants found the set screw remained securely attached within the head of the polyaxial screw.The explanted rod contains marking which indicate the set screw was fully tightened with distributed loading on the rod.There is also evidence of rod motion/slipping throughout the length of the rod.Although the wear patterns on the returned product are indicative of a construct that was properly final tightened, root cause cannot be determined.
|
The initial surgery was performed on (b)(6) 2018 from the l3-l5.After the surgery, the surgeon planned to conduct a revision surgery (for an unrelated issue) to replace some screws and extend the fixation range to s1.The revision surgery was performed on (b)(6) 2018.When the surgeon was about to remove the set screw with a set screwdriver, the poly-axial screw suddenly came off even though the set screw was not rotated.The rod as not able to be inserted between the poly-axial screw and the set screw.
|