BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold¿ lite with capio slim was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio cage did not catch the dart.When the physician tried to push again the dart into the capio cage outside the patient`s body, the dart cleanly detached from the suture.The procedure was completed with another uphold¿ lite with capio slim.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
A visual examination of the returned uphold lite with capio slim revealed that the suture on the blue dilator was broken.There was a small amount of suture attached to the dart and was returned with the device.Analysis also revealed no damage to the capio slim suture capturing device.The dart on the blue with white stripe dilator was caught in the cage of the capio slim as intended when actuated.A review of the device history record (dhr) indicated that the device met all material, assembly, and product specifications at the time of release to distribution.However, the investigation concluded that the most probable cause for this event is manufacturing process design as the design or validation of the manufacturing process was not sufficient to ensure the finished device met the intent of the design.The investigation concluded that the design of the carrier allows the fiber portion of the suture to interact with the sharp edge of the carrier, resulting in suture severing.The issue is under investigation and a correction has not yet been implemented.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold lite with capio slim was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio cage did not catch the dart.When the physician tried to push again the dart into the capio cage outside the patient`s body, the dart cleanly detached from the suture.The procedure was completed with another uphold lite with capio slim.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|