Catalog Number 82008
Device Problems Fluid/Blood Leak (1250); Overheating of Device (1437); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where an overheating insert resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a 30k fsi-sli-fg-1000 insert, the insert heats up and drips water; no injury resulted.
 
Manufacturer Narrative
The device was evaluated and found to be within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FOCUSED SPRAY SLIMLINE FITGRIP ULTRASONIC INSERT - 1000
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
MDR Report Key7378037
MDR Text Key103803461
Report Number2424472-2018-00032
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number82008
Device Lot Number17179
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/26/2018
Initial Date FDA Received03/28/2018
Supplement Dates Manufacturer Received04/10/2018
Supplement Dates FDA Received05/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-