• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.A medtronic representative went to site to test the equipment.Representative reported that a computer was replaced to resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect computer has been received by the manufacturer for evaluation.The returned computer was tested and computer passed all tests without any errors.No fault found.
 
Event Description
A medtronic representative reported that during functional endoscopic sinus surgery (fess), the navigation system unexpectedly shutdown.The procedure was completed with the use of navigation.No information was provided on delay to procedure and patient outcome.
 
Manufacturer Narrative
Correction: device manufacture dated corrected.
 
Manufacturer Narrative
Additional information: the patient demographic information was requested from (b)(6) coordinator on (b)(6) 2019 at the facility indicated the information was unknown.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system during a functional endoscopic sinus surgery (fess).It was reported that in the middle of the case the system unexpectedly shutdown.The surgery was completed with unknown amount of delay nor unknown patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
pa chia vue
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7379693
MDR Text Key103786619
Report Number1723170-2018-01363
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/07/2018
Initial Date FDA Received03/28/2018
Supplement Dates Manufacturer Received04/03/2018
02/12/2019
Supplement Dates FDA Received04/18/2018
02/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-