Model Number FUSION
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/08/2018
Event Type  malfunction  
Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.Resolved at the time of issue.
 
Event Description
A site representative reported that during functional endoscopic sinus surgery (fess), the navigation system became unresponsive to user input.Rebooting the system resolved the issue.The procedure was completed with the use of navigation.There was a delay of less than 1 hour.No known impact on patient outcome.
 
Manufacturer Narrative
Correction: device manufacture date updated to proper value.
 
Manufacturer Narrative
A software analysis was initiated to determine the probable cause of the issue through known anomaly determination.Analysis found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7383245
MDR Text Key103883133
Report Number1723170-2018-01378
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2018
Initial Date FDA Received03/29/2018
Supplement Dates Manufacturer Received04/03/2018
05/31/2018
Supplement Dates FDA Received04/19/2018
06/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age62 YR
Patient Weight111
-
-