• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT DR RF PACEMAKER; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT DR RF PACEMAKER; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2210
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/02/2018
Event Type  malfunction  
Event Description
It was reported that an asymptomatic patient presented with atrial lead noise.The noise presented as noise reversion episodes and false episodes of automodeswitch (ams).There were no sensing or impedance changes but the atrial threshold had gradually increased over the past year.On (b)(6) 2018, during a generator change for elective replacement indicator (eri), the lead was capped and replaced.Also, during the replacement procedure, it was noted that there was insulation abrasion on the atrial lead and part of the device header was worn down where the lead appeared to have been rubbing against it.The patient was stable prior, during and post procedure.
 
Manufacturer Narrative
The reported field event of header damage was not confirmed in the laboratory.The device was tested on the bench and using automated testing equipment, and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT DR RF PACEMAKER
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7383768
MDR Text Key103919613
Report Number2017865-2018-04663
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2011
Device Model NumberPM2210
Device Lot Number2885379
Other Device ID Number05414734503617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/02/2018
Initial Date FDA Received03/29/2018
Supplement Dates Manufacturer Received07/27/2018
Supplement Dates FDA Received08/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
-
-