Model Number 3058
Device Problems Unable to Obtain Readings (1516); Electromagnetic Compatibility Problem (2927)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/23/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer¿s representative (rep) regarding a patient undergoing an implant procedure for an implanted neurostimulator (ins) for gastrointestinal /pelvic floor therapy.It was reported that during the procedure today that the rep noticed an invalid reading showing "?" for the value and was unable to calculate the impedance.When "?" for the 0 pair was seen the rep increased values to 2.0 volts, 330 pulse width and then 3.0 v and 330 pulse width and moved overhead lights away and this resolved the issue.No patient symptoms were reported and no further complications were reported or are anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7385472
MDR Text Key104031947
Report Number3004209178-2018-06114
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/28/2019
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/23/2018
Initial Date FDA Received03/30/2018
Date Device Manufactured01/09/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
-
-