Model Number IBI-87008
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/14/2018
Event Type  malfunction  
Event Description
At the end of the procedure, damage was noted on the insulation of the catheter shaft.The damage likely occurred due to contact with an sl1 introducer sheath.There were no adverse consequences to the patient.
 
Manufacturer Narrative
The reported damage to the catheter shaft was confirmed.Visual inspection revealed the catheter shaft was kinked and torn.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the kink and tear in the shaft is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F
Type of Device
CATHETER, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7393725
MDR Text Key104424082
Report Number2030404-2018-00024
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K042775
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Model NumberIBI-87008
Device Lot Number6217532
Other Device ID Number05414734309929
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2018
Initial Date FDA Received04/03/2018
Supplement Dates Manufacturer Received04/16/2018
Supplement Dates FDA Received04/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-